A DEPTH-ENCODED PET DETECTOR
A DEPTH-ENCODED PET DETECTOR
复制标题
DOI:
10.1109/23.289382
复制
发表时间:
1991-04-01
影响因子:
1.8
通讯作者:
THOMPSON, CJ
中科院分区:
文献类型:
--
作者:
BARTZAKOS, P;THOMPSON, CJ
Radial blurring in high-resolution positron emission tomography (PET) can be reduced by measuring the PET photon's depth of interaction within a detector crystal. Simulations show that dividing the crystal into two equiprobable regions of interaction almost eliminates this blurring. This has been done using a black band placed around a 3 cm BGO crystal, 1 cm from the front. A PMT collects light from the rear of the crystal for energy and timing measurement. Another PMT collects light through a light guide from the top-front of the crystal to find the depth of interaction.Gamma rays interacting in the front of the crystal produce a strong, signal in the depth-sensing PMT, while interactions behind the band produce weaker (< 10%) signal. The intensity of the light recorded by the energy PMT is almost constant (within +/- 3%) for interactions at any depth in the crystal.