Evaluation of the Oxoid Aura image system for measuring zones of inhibition with the disc diffusion technique
Evaluation of the Oxoid Aura image system for measuring zones of inhibition with the disc diffusion technique
复制标题
DOI:
10.1093/jac/46.4.535
复制
发表时间:
2000-10-01
影响因子:
5.2
通讯作者:
Wise, R
中科院分区:
文献类型:
--
作者:
Andrews, JM;Boswell, FJ;Wise, R
In this study the Oxoid Aura image antibiotic sensitivity test system, used as a stand-alone device, was compared with manual zone measurement and use of a template, for the determination of sensitivities. An overall correlation coefficient of 0.99 was observed for zone diameters measured using the Aura image system and zones measured manually, when the differences between zones were within 3 mm; 5.4% of zones showed a difference in zone diameter between manual and automated measurement of >3 mm. The results obtained using the template method for interpretation were less reliable than zone measurement, with cefuroxime and ampicillin tested against Enterobacteriaceae and Acinetobacter spp. When linked to a laboratory patient database, the bar code and disc identification facilities avoided errors that were associated with manual data entry.