FastICA peel-off for ECG interference removal from surface EMG.
FastICA peel-off for ECG interference removal from surface EMG.
复制标题
FastICA 剥离用于消除表面肌电图的心电图干扰
DOI:
10.1186/s12938-016-0196-8
复制
发表时间:
2016-06-13
影响因子:
3.9
通讯作者:
Zhou P
中科院分区:
文献类型:
--
作者:
Chen M;Zhang X;Chen X;Zhu M;Li G;Zhou P
BackgroundMulti-channel recording of surface electromyographyic (EMG) signals is very likely to be contaminated by electrocardiographic (ECG) interference, specifically when the surface electrode is placed on muscles close to the heart.MethodsA novel fast independent component analysis (FastICA) based peel-off method is presented to remove ECG interference contaminating multi-channel surface EMG signals. Although demonstrating spatial variability in waveform shape, the ECG interference in different channels shares the same firing instants. Utilizing the firing information estimated from FastICA, ECG interference can be separated from surface EMG by a “peel off” processing. The performance of the method was quantified with synthetic signals by combining a series of experimentally recorded “clean” surface EMG and “pure” ECG interference.ResultsIt was demonstrated that the new method can remove ECG interference efficiently with little distortion to surface EMG amplitude and frequency. The proposed method was also validated using experimental surface EMG signals contaminated by ECG interference.ConclusionsThe proposed FastICA peel-off method can be used as a new and practical solution to eliminating ECG interference from multichannel EMG recordings.