Application of low-tube current with iterative model reconstruction on Philips Brilliance iCT Elite FHD in the accuracy of spinal QCT using a European spine phantom
Application of low-tube current with iterative model reconstruction on Philips Brilliance iCT Elite FHD in the accuracy of spinal QCT using a European spine phantom
复制标题
DOI:
10.21037/qims.2018.02.03
复制
发表时间:
2018
影响因子:
2.8
通讯作者:
Mingyue Wang
中科院分区:
文献类型:
--
作者:
Yan WU;Yaojun Jiang;Mingyue Wang
Repeated measurements of BMD with QCT in the ESP multicenter showed that BMD changes in L1–L3 are the least varied at IMR [2] algorithm. It is recommended to scan at 120 kV with 20 mAs combined with IMR [2] algorithm. In this way, the BMD of spine by QCT could be accurately measured, while radiation dosage significantly reduced and imaging quality improved at the same time.