IDENTIFICATION OF POSITIVE REACTANT IONS OBSERVED FOR NITROGEN CARRIER GAS IN PLASMA CHROMATOGRAPH MOBILITY STUDIES
IDENTIFICATION OF POSITIVE REACTANT IONS OBSERVED FOR NITROGEN CARRIER GAS IN PLASMA CHROMATOGRAPH MOBILITY STUDIES
复制标题
DOI:
10.1021/ac60362a029
复制
发表时间:
1975-01-01
影响因子:
7.4
通讯作者:
HORNING, EC
中科院分区:
文献类型:
--
作者:
CARROLL, DI;DZIDIC, I;HORNING, EC
A Plasma Chromatograph-mass spectrometer (PC-MS) combined Instrument was used to define the Ion species re-sponsible for the three major peaks observed In positive Ion mobility studies of nitrogen carrier gas at160 C and 7 ppm water concentration. These species are: l, NH4+ with about 7% of NH4+(H20); II, NO+ with about 10% of NO+(H20); III, H+(H20) 2 and H+(H20) 3 In about 7: 3 ratio, for ion peaks showing reduced mobilities of 3.00, 2.62, and 2.32 cm2 v-1 sec-1, respectively. Literature Identifications basedupon presumed equivalence of mobility and mass measurements are in error.The early mass spectrometric studies of Shahin (1), employing a corona discharge in air at atmospheric pressure, indicate thatH+(H20)„ions should be the dominant species in positive ion mobility spectra observed with nitro-gen carrier gas in the Plasma Chromatograph. This was confirmed using a Plasma Chromatograph-mass spectrom-eter (PC-MS) combined instrument at Franklin GNO Corporation (2). The ions NO+ and NO+(H20) were also found to be present (3).