我们开发了一种新的读出系统的近红外探测器VIRGO-2K(2kx 2k HgCdTe阵列)安装在光学-红外同步相机,HONIR,为1.5米金田望远镜在东广岛天文台。此开发的主要目标是通过探测器的16个输出读出模式在约1秒内读出一帧,以减少每次曝光的开销时间。该系统是基于一个CCD控制器,Kiso阵列控制器(KAC)。我们重新设计了KAC的模拟部分,以适应VIRGO-2K。采用全差分输入电路和三阶Bessel低通滤波器来降低噪声,采用恒流系统来提高检波器的线性度。我们将贝塞尔低通滤波器的截止频率设置为读出时钟速率(120 kHz)。我们还根据VIRGO-2K的数据表将恒定电流设置为200 μA。我们在室温下测试了新的读出系统,并确认低通滤波器按设计运作良好。在VIRGO-2K的典型输出电压范围(3.2-4.4 V)内,恒流系统的电流水平波动小于2%。我们测量了由新读出系统(连接到冷却多路复用器)引起的读出噪声,发现其为30-40 μV rms,与VIRGO-2K的典型读出噪声(37 μV rms)相当或略高。
We developed a new readout system for the near-infrared detector VIRGO-2K (2kx2k HgCdTe array) installed in the optical-infrared simultaneous camera, HONIR, for the 1.5 m Kanata telescope at Higashi-Hiroshima observatory. The main goal of this development is to read out one frame within ~ 1 second through 16 output readout mode of the detector, in order to reduce the overhead time per exposure. The system is based on a CCD controller, Kiso Array Controller (KAC). We redesigned the analog part of KAC to fit VIRGO-2K. We employed a fully differential input circuit and a third order Bessel low-pass filter for noise reduction and a constant current system to improve the linearity of the detector. We set the cutoff frequency of the Bessel low-pass filter at the readout clock rate (120 kHz). We also set the constant current at 200 μA according to the data sheet of VIRGO-2K. We tested the new readout system at room temperature and confirmed that the low-pass filter works well as designed. The fluctuation of the current level of the constant current system is less than 2% for the typical output voltage range of VIRGO-2K (3.2-4.4 V). We measured the readout noise caused by the new readout system (connected to cooled multiplexer) and found that it is 30-40 μV rms, being comparable to or slightly higher than the typical readout noise of VIRGO-2K, ∼ 37 μV rms.
光学和近红外多用途仪器 HONIR
DOI: --
发表时间: 2012
期刊:
影响因子: --
作者:
Sakimoto;K.;Kawabata;K.~S.;Yoshida;M.;et al.
通讯作者: et al.