Development of a new readout system for the near-infrared detector of HONIR
Development of a new readout system for the near-infrared detector of HONIR
复制标题
HONIR近红外探测器新型读出系统的开发
DOI:
10.1117/12.2055388
复制
发表时间:
2014
期刊:
影响因子:
--
通讯作者:
Y. Hashiba
中科院分区:
文献类型:
--
作者:
T. Ui;S. Sako;T. Yamashita;H. Akitaya;K. Kawabata;H. Nakaya;Y. Moritani;R. Itoh;K. Takaki;T. Urano;I. Ueno;T. Ohsugi;Michitoshi Yoshida;H. Nakao;Y. Hashiba
We developed a new readout system for the near-infrared detector VIRGO-2K (2kx2k HgCdTe array) installed in the optical-infrared simultaneous camera, HONIR, for the 1.5 m Kanata telescope at Higashi-Hiroshima observatory. The main goal of this development is to read out one frame within ~ 1 second through 16 output readout mode of the detector, in order to reduce the overhead time per exposure. The system is based on a CCD controller, Kiso Array Controller (KAC). We redesigned the analog part of KAC to fit VIRGO-2K. We employed a fully differential input circuit and a third order Bessel low-pass filter for noise reduction and a constant current system to improve the linearity of the detector. We set the cutoff frequency of the Bessel low-pass filter at the readout clock rate (120 kHz). We also set the constant current at 200 μA according to the data sheet of VIRGO-2K. We tested the new readout system at room temperature and confirmed that the low-pass filter works well as designed. The fluctuation of the current level of the constant current system is less than 2% for the typical output voltage range of VIRGO-2K (3.2-4.4 V). We measured the readout noise caused by the new readout system (connected to cooled multiplexer) and found that it is 30-40 μV rms, being comparable to or slightly higher than the typical readout noise of VIRGO-2K, ∼ 37 μV rms.
DOI:
--
发表时间:
2012
期刊:
影响因子:
--
作者:
Sakimoto;K.;Kawabata;K.~S.;Yoshida;M.;et al.
通讯作者:
et al.