Viscosity Measurements in Liquid Using Transversal Effects of Piezoceramic Disk-Type PZT Transducer in the Frequency Range of 70–430 kHz
Viscosity Measurements in Liquid Using Transversal Effects of Piezoceramic Disk-Type PZT Transducer in the Frequency Range of 70–430 kHz
复制标题
在 70–430 kHz 频率范围内利用压电陶瓷盘式 PZT 传感器的横向效应测量液体粘度
DOI:
10.1143/jjap.40.3654
复制
发表时间:
2001
期刊:
影响因子:
--
通讯作者:
K. Imano
中科院分区:
文献类型:
--
作者:
S. Momozawa;K. Imano
Transversal effects of a piezoceramic disk-type Pb(Zr,Ti)O3 (PZT) transducer are newly introduced in the measurement of dynamic viscosity of liquid. The fundamental resonance (75 kHz) and the higher harmonic resonant frequencies (195, 310, 424 kHz) of the transducer are used in the experiments. The resonant resistance Rr and the resonant frequency shift Δfr of transducer are measured with a water-glycerin mixtures (viscosity ηL:3~490 mPas) and silicone oil (ηL:10 ~1000 mPas) as the sample liquids. From the experimental results, it is gathered that 1) the resonant resistance Rr has a linear relationship to ηL1/2, 2) the resonant frequency shift Δfr does not show a linearity in the high viscosity range larger than 100 mPas. The viscosity measurement in the frequency range of 75~424 kHz, which has not been explored previously, was successfully accomplished by our method.