Amperometric determination of phosphatidyl choline in serum with use of immobilized phospholipase d and choline oxidase
Amperometric determination of phosphatidyl choline in serum with use of immobilized phospholipase d and choline oxidase
复制标题
固定化磷脂酶 d 和胆碱氧化酶安培法测定血清中磷脂酰胆碱
DOI:
10.1016/s0003-2670(01)85007-8
复制
发表时间:
1979
影响因子:
6.2
通讯作者:
Shuichi Suzuki
中科院分区:
文献类型:
--
作者:
I. Karube;K. Hara;I. Satoh;Shuichi Suzuki
Phospholipase D (EC 3.1.4.4.) and choline oxidase (EC 1.1.99.1.) are immobilized together on a hydrophobic agarose gel and used to convert the phospholipid to betaine and hydrogen peroxide, which is measured amperometrically at + 0.60 V vs. SCE. The response time of the sensor is 2 min, and the calibration curve for 0–3 g l-1of phosphatidyl choline is linear. Different methods of insolubilizing the enzymes are compared.