Measurement of thoron gas in the environment using a Lucas scintillation cell
Measurement of thoron gas in the environment using a Lucas scintillation cell
复制标题
DOI:
10.1088/0952-4746/30/3/013
复制
发表时间:
2010-09
影响因子:
1.5
通讯作者:
Lei Zhang;Jian Wu;Q. Guo;W. Zhuo
中科院分区:
文献类型:
--
作者:
Lei Zhang;Jian Wu;Q. Guo;W. Zhuo
A simple and fast method for measuring the concentration of 220Rn in the environment was developed based on the AB-5 portable radon measuring device. First, background counts were measured with a Lucas scintillation cell (LSC), then air sampling measurement was started immediately and lasted for 1 min; the 220Rn concentration could be calculated from the counts before and after sampling as well as the theoretical detection efficiency. Results of theoretical calculations showed that in a pure 220Rn environment the lower detection limit of the LSC would be some 65 Bq m−3 (with a confidence level of 70%). The experimental results showed that compared with a RAD7 monitor in pure 220Rn or mixed 222Rn/220Rn environments the deviations were less than ± 10%. This method can also be used for quantitative measurement of 220Rn concentration.