Localization of Transcranial Targets for Photoacoustic-Guided Endonasal Surgeries.

Localization of Transcranial Targets for Photoacoustic-Guided Endonasal Surgeries.
复制标题

DOI:
10.1016/j.pacs.2015.05.002
复制
发表时间:
2015-06
期刊:
影响因子:
7.9
通讯作者:
Boctor EM
Boctor EM
中科院分区:
工程技术1区
文献类型:
--
作者:
Lediju Bell MA;Ostrowski AK;Li K;Kazanzides P;Boctor EM

文献摘要

被引文献

相似文献

Neurosurgeries to remove pituitary tumors using the endonasal, transsphenoidal approach often incur the risk of patient death caused by injury to the carotid arteries hidden by surrounding sphenoid bone. To avoid this risk, we propose intraoperative photoacoustic vessel visualization with an optical fiber attached to the surgical tool and an external ultrasound transducer placed on the temple. Vessel detection accuracy is limited by acoustic propagation properties, which were investigated with k-Wave simulations. In a two-layer model of temporal bone (3200 m/s sound speed, 1-4 mm thickness) and surrounding tissues, the localization error was ≤2 mm in the tranducer's axial dimension, while temporal bone curvature further degraded target localization. Phantom experiments revealed that multiple image targets (e.g. sphenoid bone and vessels) can be visualized, particularly with coherence-based beamforming, to determine tool-to-vessel proximity despite expected localization errors. In addition, the potential flexibility of the fiber position relative to the transducer and vessel was elucidated.