Stimulus and transducer effects on threshold.
Stimulus and transducer effects on threshold.
复制标题
DOI:
10.3109/14992027.2014.979300
复制
发表时间:
2015-02
影响因子:
2.7
通讯作者:
Stephenson MR
中科院分区:
文献类型:
--
作者:
Flamme GA;Geda K;McGregor KD;Wyllys K;Deiters KK;Murphy WJ;Stephenson MR
This study examined differences in thresholds obtained under Sennheiser HDA200 circumaural earphones using pure tone, equivalent rectangular noise bands, and 1/3 octave noise bands relative to thresholds obtained using Telephonics TDH-39P supra-aural earphones. Thresholds were obtained via each transducer and stimulus condition six times within a 10-day period. Forty-nine adults were selected from a prior study to represent low, moderate, and high threshold reliability. The results suggested that (1) only small adjustments were needed to reach equivalent TDH-39P thresholds, (2) pure-tone thresholds obtained with HDA200 circumaural earphones had reliability equal to or better than those obtained using TDH-39P earphones, (3) the reliability of noise-band thresholds improved with broader stimulus bandwidth and was either equal to or better than pure-tone thresholds, and (4) frequency-specificity declined with stimulus bandwidths greater than one Equivalent Rectangular Band, which could complicate early detection of hearing changes that occur within a narrow frequency range. These data suggest that circumaural earphones such as the HDA200 headphones provide better reliability for audiometric testing as compared to the TDH-39P earphones. These data support the use of noise bands, preferably ERB noises, as stimuli for audiometric monitoring.