Clinical experience with the Fieldmaster perimeter.

Clinical experience with the Fieldmaster perimeter.
复制标题

Fieldmaster 周界的临床经验。

DOI:
10.1016/0002-9394(82)90420-2
复制
发表时间:
1982
影响因子:
4.2
通讯作者:
R. Drews
R. Drews
中科院分区:
医学1区
文献类型:
--
作者:
J. C. Bobrow;R. Drews

文献摘要

被引文献

相似文献

在临床实践环境中,使用 Fieldmaster Model 200 对连续 167 名患者进行的视野检查表明,Fieldmaster 测试引出了 Goldmann 动力学视野检查中发现的 95% 的缺陷,几乎没有假阳性或假阴性结果;相对不熟练的人员也能够很好地进行 Fieldmaster 测试;随着我们的人员和患者对检测的熟悉,检查不完整率下降到了 2% 以下。
Perimetry performed on a Fieldmaster Model 200 on a consecutive series of 167 patients in a clinical practice setting demonstrated that Fieldmaster testing elicited 95% of defects found on Goldmann kinetic perimetry with very few false-positive or false-negative results; relatively unskilled personnel were able to conduct Fieldmaster testing well; and as our personnel and patients became familiar with the test, the rate of incomplete examinations declined to less than 2%.