Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beam
Absorbed dose estimation using LET dependence in glow curve of thermoluminescent phosphor Li3B7O12:Cu in therapeutic carbon beam
复制标题
利用治疗碳束中热释光磷光体 Li3B7O12:Cu 的辉光曲线中的 LET 依赖性估算吸收剂量
DOI:
10.1080/00223131.2016.1181010
复制
发表时间:
2016
影响因子:
1.2
通讯作者:
Kiyomitsu Shinsho
中科院分区:
文献类型:
--
作者:
Weishan Chang;Yusuke Koba;Shigekazu Fukuda;Genichiro Wakabayashi;Hidetoshi Saitoh;Kiyomitsu Shinsho
A high temperature ratio (HTR) method has been proposed to correct the linear energy transfer (LET) dependence of thermoluminescent (TL) efficiency. To realize the use of the slab-type thermoluminescence detector (TLD) that based on the phosphor Li3B7O12:Cu for heavy charged particle beam, the HTR method has been considered. To improve the reproducibility of HTR, the slow heating rate method is introduced in this report and the coefficient variations of HTR decreased from 10%–20% to 8%. The relation between TL-efficiency, HTR, and LET for Li3B7O12:Cu was manifested and the TL-efficiency as a function of HTR was derived in an attempt to measure the absorbed dose without LET information. The feasibility of the HTR method in therapeutic carbon beams was evaluated by comparing the dose estimated by Li3B7O12:Cu and by an ionization chamber. The accuracy of dose estimation in carbon beams was improved by using the HTR method, but there is room for further improvement. The use of Li3B7O12:Cu in heavy charged particle beams can be materialized with further improvement of HTR sensitivity.