In vivo measurement of anodic polarization of orthopaedic implant alloys: comparative study of in vivo and in vitro experiments.

In vivo measurement of anodic polarization of orthopaedic implant alloys: comparative study of in vivo and in vitro experiments.
复制标题

骨科植入合金阳极极化的体内测量:体内和体外实验的比较研究。

DOI:
10.1016/0142-9612(89)90134-8
复制
发表时间:
1989
期刊:
影响因子:
14
通讯作者:
M. Oka
M. Oka
中科院分区:
工程技术1区
文献类型:
--
作者:
Y. Nakayama;T. Yamamuro;Y. Kotoura;M. Oka

文献摘要

被引文献

相似文献

应用一种新的单片电极系统对几种骨科植入物用合金的阳极极化进行了活体和体外测量。每次实验结束后,用体视显微镜仔细观察被测表面。这证实了凹坑的形成,并排除了其他突然增加的阳极电流的可能性,如缝隙腐蚀和有机物的氧化或还原。结果重现性好。虽然在两种条件下得到的结果有一定差异,但合金的耐蚀性顺序是一致的。
A new system of single piece electrodes was applied toIn vivoandin vitromeasurements of anodic polarization of several alloys used in orthopaedic implants. The tested surface was scrutinized with a stereomicroscope after each experiment. This confirmed pit formation and excluded the possibility of other sudden increase of anodic current such as crevice corrosion and oxidation or reduction of organic substances. The results were reproducible. Although there were certain differences between the results obtained in the two conditions, the order of corrosion resistances of alloys was consistent.