Sensitive and real-time detection of IgG using interferometric reflecting imaging sensor system
Sensitive and real-time detection of IgG using interferometric reflecting imaging sensor system
复制标题
DOI:
10.1016/j.bios.2021.113961
复制
发表时间:
2022-01-10
影响因子:
12.6
通讯作者:
Denizli, Adil
中科院分区:
文献类型:
--
作者:
Bakhshpour, Monireh;Chiodi, Elisa;Denizli, Adil
Considering the limitations of well-known traditional detection techniques, innovative research studies have focused on the development of new sensors to offer label-free, highly sensitive, real-time, low-cost, and rapid detection for biomolecular interactions. In this study, we demonstrate immunoglobulin G (IgG) detection in aqueous solutions by using real-time and label-free kinetic measurements of the Interferometric Reflectance Imaging Sensor (IRIS) system. By performing kinetic characterization experiments, the sensor's performance is comprehensively evaluated and a high correlation coefficient value (>0.94) is obtained in the IgG concentration range of 1-50 mu g/mL with a low detection limit (0.25 mu g/mL or 1.67 nM). Moreover, the highly sensitive imaging system ensures accurate quantification and reliable validation of recorded binding events, offering new perspectives in terms of direct biomarker detection for clinical applications.