QUANTIFIED DEEP TENDON REFLEX DEVICE, SECOND GENERATION

QUANTIFIED DEEP TENDON REFLEX DEVICE, SECOND GENERATION
复制标题

量化深肌腱反射装置,第二代

DOI:
10.1142/s0219519408002462
复制
发表时间:
2008
影响因子:
0.8
通讯作者:
R. Jafari
R. Jafari
中科院分区:
工程技术4区
文献类型:
--
作者:
R. LeMoyne;F. Dabiri;R. Jafari

文献摘要

被引文献

相似文献

The deep tendon reflex is a fundamental aspect of neurological examinations. The severity of and degree of recovery from a traumatic brain injury can be assessed by the myotatic stretch reflex. A hyperactive reflex response is correlated with spasticity, which can also be correlated with the degree of damage to the supraspinal input, in essence assessing the severity of traumatic brain injury. The myotatic stretch reflex is clinically evaluated by the National Institute of Neurological Disorders and Stroke (NINDS) reflex scale (0–4); however, this scale lacks temporal data and may also vary in interpretation. The solution is a fully quantified evaluation system of the myotatic stretch reflex, whereby a patellar hammer's force input is based on original potential energy and a microelectromechanical system (MEMS) accelerometer quantifies the output. The MEMS accelerometer is attached to a set anchor point near the ankle. The reflex amplitude is based on the maximum acceleration of the reflex response. The quantified data collected from MEMS accelerometers are transmitted by a portable computer (i.e. a Pocket PC). This paper describes a device that quantitatively evaluates the reflex response using accelerometers and that demonstrates precision for reproducibility.