A method was established for simultaneous determination of perfluorooctane sulfonate and perfluorooctanoic acid isomers in environmental sample by using high performance liquid chromatography-tandem mass spectrometry coupled with solid phase extraction (SPE-HPLC-MS/MS). The WAX SPE cartridge was used for samples' enrichment and clean-up, then followed by determination by HPLC-MS/MS in negative electrospray ionization and multiple reaction monitoring (MRM) mode. The separation was achieved with an Ascentis (R) Express F5 PFP column with mobile phases of A, methanol and B, 60 mmol . L-1 NH4OH/20 mmol . L-1 formic acid aqueous solution. The established method was used for the analyses of water, sediment and shellfish samples. The method detection limits(S/N = 3) were in the range of 0.05-1.1 ng . L-1, 0.025-0.56 ng . g(-1) and 0.025-0.56 ng . g(-1), respectively. Satisfactory results were obtained with the recoveries of PFASs isomers in the range of 90. 8%-127%, 74. 0%-124% and 76. 7%-113%, and the relative standard deviations (RSD, n = 5) in the range of 5.9%-17.2%, 2.5%-17.0% and 3.30%-11.5%, respectively.