Half-Width Integral Method for Gas Concentration Measuring in Tunable Diode Laser Absorption Spectroscopy
Half-Width Integral Method for Gas Concentration Measuring in Tunable Diode Laser Absorption Spectroscopy
复制标题
DOI:
10.1080/00387010.2012.697090
复制
发表时间:
2013-04
影响因子:
1.7
通讯作者:
Ding Yanjun;Li Xiaohang;Peng Zhimin;Che Lu
中科院分区:
文献类型:
--
作者:
Ding Yanjun;Li Xiaohang;Peng Zhimin;Che Lu
ABSTRACT We analyze the influence of random errors and absorption shape fitting errors on gas concentration measurement in tunable diode laser absorption spectroscopy. We then propose a new data processing method according to the characteristic of random errors and fitting absorption shape. This method only uses the integral value within the half width at half maximum of the fitting absorption shape to determine the gas concentration in actual measurements. Meanwhile, the O2 transition at 13,150.197 cm−1 is selected to measure the O2 concentration using this method. Compared with the traditional method, the method established in this paper can accurately measure O2 concentration.