Escherichia coli O157:H7 detection limit of millimeter-sized PZT cantilever sensors is 700 Cells/mL
Escherichia coli O157:H7 detection limit of millimeter-sized PZT cantilever sensors is 700 Cells/mL
复制标题
DOI:
10.2116/analsci.21.355
复制
发表时间:
2005-04-01
影响因子:
1.6
通讯作者:
Mutharasan, R
中科院分区:
文献类型:
--
作者:
Campbell, GA;Mutharasan, R
A composite self-excited millimeter-sized lead zirconate titanate (PZT) glass cantilever (2 mm x 1.8 mm; sensing area of 6 mm(2)) was fabricated for the detection of Escherichia coli (E. coli) O157:H7. The fundamental and second mode resonance in air was 10.95 +/- 0.05 kHz and 43.45 +/- 0.05 kHz, respectively. Affinity purified monoclonal antibody (anti-E. coli O157:H7) specific to the pathogen E. coli O157:H7 was immobilized at the cantilever glass tip, and then immersed in liquid containing the pathogen (70 to 7 x 107 cells/mL). The resonant frequency showed a reduction and reached a steady state shift of 0 +/- 5, 46 +/- 5, 260 +/- 5, and 1010 5 Hz corresponding to 0, 700, 7000, and 7 x 10(7) cells/mL. From the experiments conducted, the detection limit of the sensor was 700 cells/mL.