An integrated AC electrokinetic pump in a microfluidic loop for fast and tunable flow control
An integrated AC electrokinetic pump in a microfluidic loop for fast and tunable flow control
复制标题
DOI:
10.1039/b408382m
复制
发表时间:
2004-01-01
期刊:
影响因子:
4.2
通讯作者:
Ajdari, A
中科院分区:
文献类型:
--
作者:
Studer, V;Pépin, A;Ajdari, A
We have built a dedicated lab on a chip to study the performance of an integrated electrokinetic micropump, driven by a low voltage AC signal. This micropump consists of an array of interdigitated electrodes and is here integrated in a microfluidic loop. We demonstrate that this device can pump continuously and reproducibly electrolyte solutions of low to moderate ionic strength. The pumping speed reaches up to 500 mum s(-1) in 20 mum deep and 100 mm wide channels with a driving signal in the 1-10 kHz range and an amplitude of only a few volts. In addition, we have observed an interesting reversal of the pumping direction at higher frequencies (50-100 kHz). Our device permits a systematic and automated exploration of the influence of the ionic strength thanks to an integrated micromixer.