Comparison Between Diffusion-Weighted MRI (DW-MRI) at 1.5 and 3 Tesla: A Phantom Study
Comparison Between Diffusion-Weighted MRI (DW-MRI) at 1.5 and 3 Tesla: A Phantom Study
复制标题
DOI:
10.1002/jmri.24397
复制
发表时间:
2014-09-01
影响因子:
4.4
通讯作者:
Aboagye, Eric O.
中科院分区:
文献类型:
--
作者:
Lavdas, Ioannis;Miquel, Marc E.;Aboagye, Eric O.
Purpose: To compare DW-MRI between 1.5 and 3 Tesla (T) in terms of image quality, apparent diffusion coefficient (ADC), reproducibility, lesion-to-background contrast and signal-to-noise ratio (SNR), using a test object.Materials and Methods: A spherical diffusion phantom was used for qualitatively assessing image quality and performing quantitative measurements between the two field strengths.Results: Distortions and signal losses degraded image quality at 3T even when the protocols were optimized for minimum TE. The ADC, in the majority of the phantom compartments, was significantly different between 1.5T and 3T (P < 0.009), while the average coefficient of variation, excluding the phantom compartments affected by artifacts, was