Changes in proximal femoral strain after insertion of uncemented standard and customised femoral stems

Changes in proximal femoral strain after insertion of uncemented standard and customised femoral stems
复制标题

DOI:
10.1302/0301-620x.83b6.9726
复制
发表时间:
2001-08-01
影响因子:
--
通讯作者:
Husby, OS
Husby, OS
中科院分区:
其他
文献类型:
--
作者:
Aamodt, A;Lund-Larsen, J;Husby, OS

文献摘要

被引文献

相似文献

We have compared the changes in the pattern of the principal strains in the proximal femur after insertion of eight uncemented anatomical stems and eight customised stems in human cadaver femora. During testing we aimed to reproduce the physiological loads on the proximal femur and to simulate single-leg stance and stair-climbing.The strains in the intact femora were measured and there were no significant differences in principal tensile and compressive strains in the left and right femora of each pair. The two types of femoral stem were then inserted randomly into the left or right femora and the cortical strains were again measured. Both induced significant stress shielding in the proximal part of the metaphysis, but the deviation from the physiological strains was most pronounced after insertion of the anatomical stems.The principal compressive strain at the calcar was reduced by 90% for the anatomical stems and 67% for the customised stems. Medially, at the level of the lesser trochanter, the corresponding figures were 59% and 21%. The anatomical stems induced more stress concentration on the anterior aspect of the femur than did the customised stems. They also increased the hoop strains in the proximomedial femur. Our study shows a consistently more physiological pattern of strain in the proximal femur after insertion of customised stems compared with standard, anatomical stems.