Compact polytetrafluoroethylene assembly-type capillary electrophoresis with chemiluminescence detection.
Compact polytetrafluoroethylene assembly-type capillary electrophoresis with chemiluminescence detection.
复制标题
DOI:
10.1016/j.chroma.2006.06.101
复制
发表时间:
2006-08
期刊:
影响因子:
--
通讯作者:
K. Tsukagoshi;Shingo Ishida;Yuichi Oda;K. Noda;R. Nakajima
中科院分区:
文献类型:
--
作者:
K. Tsukagoshi;Shingo Ishida;Yuichi Oda;K. Noda;R. Nakajima
We have developed a compact polytetrafluoroethylene (PTFE) assembly-type capillary electrophoresis with chemiluminescence (CL) detection system. Luminol–microperoxidase–hydrogen peroxide chemiluminescence reaction was adopted. The device is rectangular in shape (60mm×40mm×30mm) and includes three reservoirs (sample, migration buffer, and detection reservoirs) with electrodes. The detection reservoir includes an optical fiber to transport light at the capillary tip to a photomultiplier tube. Isoluminol isothiocyanate (ILITC) was analyzed as a model using this device with fused-silica or polytetrafluoroethylene capillary tubes 10cm in length. We also used the sample reservoir as a reactor for an immune reaction between anti-human serum albumin immobilized on glass beads and isoluminol isothiocyanate-labeled human serum albumin. The present polytetrafluoroethylene assembly with the capillary tube was useful as a palm-sized analysis device for separation and detection, as well as a reactor.