EVALUATION OF A FIBER OPTIC GLOVE FOR SEMI-AUTOMATED GONIOMETRIC MEASUREMENTS

EVALUATION OF A FIBER OPTIC GLOVE FOR SEMI-AUTOMATED GONIOMETRIC MEASUREMENTS
复制标题

DOI:
10.1682/jrrd.1990.10.0411
复制
发表时间:
1990-01-01
影响因子:
--
通讯作者:
ROSEN J M
ROSEN J M
中科院分区:
其他
文献类型:
--
作者:
WISE S;GARDNER W;ROSEN J M

文献摘要

被引文献

相似文献

正常受试者被用来评估半自动测角测量的光纤仪器手套。手套通过测量关节弯曲的量来电子记录并将手和手指的位置传输到主机。手套通过了一系列的活动范围(ROM)的测试与五个主题。在重复标准化运动过程中,比较五个手指的掌骨(MP)和近端指间关节(PIP)角度,以评估手套的可重复性。结果显示,总体误差为5.6度,而手动测量的误差在5到8度之间。进行了额外的测试,以确定可能导致整体误差的因素,如配合、握力和手腕运动。手套应具有作为半自动测角测量装置的手部评估的某些方面的适用性。
Normal subjects were used to evaluate a fiber optic instrumented glove for semi-automated goniometric measurement. The glove electronically records and transmits hand and finger position to a host computer by measuring the amount of joint flexion. The glove was put through a series of range-of-motion (ROM) tests with five subjects. Metacarpal (MP) and proximal interphalangeal (PIP) joint angles of the five digits were compared during repetitive standardized motions to evaluate the glove's repeatability. The results showed an overall error of 5.6 degrees, as compared to an error of between 5 and 8 degrees with manual measurement. Additional tests were done to determine factors such as fit, grip force, and wrist motion that may contribute to the overall error. The glove should have applicability to some aspects of hand evaluation as a semi-automated goniometric measurement device.