Ultra-sensitive cavity ring-down spectroscopy in the mid-infrared spectral region.
Ultra-sensitive cavity ring-down spectroscopy in the mid-infrared spectral region.
复制标题
DOI:
10.1364/ol.41.001612
复制
发表时间:
2016-04-01
期刊:
影响因子:
3.6
通讯作者:
Hodges JT
中科院分区:
文献类型:
--
作者:
Long DA;Fleisher AJ;Liu Q;Hodges JT
We describe an ultra-sensitive cavity ring-down spectrometer which operates in the mid-infrared spectral region near 4.5 µm. With this instrument a noise-equivalent absorption coefficient of 2.6×10−11 cm−1 Hz−1/2 was demonstrated with less than 150 nW of optical power incident on the photodetector. Quantum noise was observed in the individual ring-down decay events, leading to quantum-noise-limited short-time performance. We believe that this spectrometer’s combination of high sensitivity and robustness make it well suited for measurements of ultra-trace gas species as well as applications in optics and fundamental physics.