Determination of the response function for two personal neutron dosemeter designs based on PADC.
Determination of the response function for two personal neutron dosemeter designs based on PADC.
复制标题
基于 PADC 确定两种个人中子剂量计设计的响应函数。
DOI:
10.1093/rpd/nct256
复制
发表时间:
2014
影响因子:
1
通讯作者:
M. Boschung
中科院分区:
文献类型:
--
作者:
S. Mayer;F. Assenmacher;M. Boschung
Since 1998 neutron dosimetry based on PADC (poly allyl diglycol carbonate) is done with a so-called original Paul Scherrer Institute (PSI) design at PSI. The original design (i.e. holder) was later changed. Both designs are optimised for use in workplaces around high-energy accelerators, where the neutron energy spectra are dominated by fast neutrons ranging up to some 100 MeV. In addition to the change of the dosemeter design a new evaluation method based on a microscope scanning technique has been introduced and the etching conditions have been optimised. In the present work, the responses obtained with the original and the new dosemeter designs are compared for fields of radionuclide sources and monoenergetic reference fields using the new evaluation method. The response curves in terms of the personal dose equivalent for normally incident neutrons were built as functions of the incident neutron energy.