Response of germanium detectors for high-energy γ-rays by 27Al(p, γ)28Si at Ep = 992keV
Response of germanium detectors for high-energy γ-rays by 27Al(p, γ)28Si at Ep = 992keV
复制标题
27Al(p, γ)28Si 锗探测器在 Ep = 992keV 时对高能 γ 射线的响应
DOI:
10.1093/ptep/ptad055
复制
发表时间:
2023
影响因子:
3.5
通讯作者:
Sakurai Hiroyoshi
中科院分区:
文献类型:
--
作者:
Mizuno Rurie;Niikura Megumi;Ikeda Tokihiro;Matsuzaki Teiichiro;Go Shintaro;Saito Takeshi Y;Michimasa Shin’ichiro;Sakurai Hiroyoshi
The performance of germanium detectors for high-energy γ-rays was evaluated using a 992-keV resonance in the27Al(p, γ)28Si reaction. The measurement was conducted at the RIKEN tandem accelerator. The energy of the excited state from the resonance was evaluated as 12540.7(2) keV. Using newly evaluated excitation energy, an energy calibration function and a photo peak efficiency of Ge detectors up to 10.8-MeV photons were deduced. Energy accuracy is achieved at 0.3 keV for the overall energy region. This reaction provides reliable energy and efficiency standards for high-energy γ-rays.