How to use high-sensitivity cardiac troponins in acute cardiac care
How to use high-sensitivity cardiac troponins in acute cardiac care
复制标题
DOI:
10.1093/eurheartj/ehs154
复制
发表时间:
2012-09-01
影响因子:
39.3
通讯作者:
Jaffe, Allan S.
中科院分区:
文献类型:
--
作者:
Thygesen, Kristian;Mair, Johannes;Jaffe, Allan S.
Recommendations for the use of cardiac troponin (cTn) measurement in acute cardiac care have recently been published. 1 Subsequently, a high-sensitivity (hs) cTn T assay was introduced into routine clinical practice. 2 This assay, as others, called highly sensitive, permits measurement of cTn concentrations in significant numbers of apparently illness-free individuals. These assays can measure cTn in the single digit range of nanograms per litre (¼picograms per millilitre) and some research assays even allow detection of concentrations, 1 ng/L. 2–4 Thus, they provide a more precise calculation of the 99th percentile of cTn concentration in reference subjects (the recommended upper reference limit [URL]). These assays measure the URL with a coefficient of variation (CV), 10%. 2–4 The high precision of hs-cTn assays increases their ability to determine small differences in cTn over time. Many assays currently in use have a CV. 10% at the 99th percentile URL limiting that ability. 5–7 However, the less precise cTn assays do not cause clinically relevant false-positive diagnosis of acute myocardial infarction (AMI) and a CV, 20% at the 99th percentile URL is still considered acceptable. 8