A comparison of the Coulter STKS, Coulter S+IV, and manual analysis of white blood cell differential counts in a human immunodeficiency virus-infected population.
A comparison of the Coulter STKS, Coulter S+IV, and manual analysis of white blood cell differential counts in a human immunodeficiency virus-infected population.
复制标题
Coulter STKS、Coulter S IV 和人类免疫缺陷病毒感染人群白细胞分类计数手动分析的比较。
DOI:
10.1093/ajcp/100.6.611
复制
发表时间:
1993
影响因子:
3.5
通讯作者:
Steigbigel,RT
中科院分区:
文献类型:
--
作者:
Cohen,AJ;Peerschke,EI;Steigbigel,RT
The white blood cell differential counts obtained by manual analysis and those obtained by automated analysis using Coulter STKS and Coulter S+IV cell counters (Coulter Electronics, Hialeah, FL) in a human immunodeficiency virus (HIV)-infected population were compared. Linear correlation analyses showed the STKS to be less accurate than the S+IV in determinations of the lymphocyte and granulocyte count. This finding was confirmed by difference analyses and by determining the number of measurements that fell outside of 95% confidence limits. Similar analyses were also done on patients not known to be HIV infected, and there was no disparity found between the STKS and S+IV in that group. In the HIV-infected population, the discrepancy between the manual and automated granulocyte counts performed on the STKS, but not on the S+IV, increased with increasing red blood cell mean corpuscular volume. Physicians caring for, and investigators performing studies on, HIV-infected patients should be aware of the potential for falsely low granulocyte numbers when using the differential from the STKS.