A Transmission Line Model for the Evaluation of MRI RF-Induced Fields on Active Implantable Medical Devices
A Transmission Line Model for the Evaluation of MRI RF-Induced Fields on Active Implantable Medical Devices
复制标题
用于评估有源植入医疗设备上 MRI 射频感应场的传输线模型
DOI:
10.1109/tmtt.2018.2851975
复制
发表时间:
2018
影响因子:
4.3
通讯作者:
Ji Chen
中科院分区:
文献类型:
--
作者:
Jingshen Liu;Jianfeng Zheng;Qingyan Wang;W. Kainz;Ji Chen
In this paper, a transmission line model is applied to evaluate magnetic resonance imaging (MRI) radio frequency (RF)-induced fields on active implantable medical devices (AIMDs). The influences of the AIMD lead body, the lead tip, and the implantable pulse generator (IPG) are analyzed based on this model. The transmission line model, together with the transfer function method, forms a systematic approach for analyzing the MRI RF safety problems. It can be used to understand the mechanisms and predict the MRI RF safety risks for various IPGs and leads. Both numerical simulation and measurement results demonstrate the accuracy and the efficiency of this method.