HIGH-RESOLUTION GAS-CHROMATOGRAPHY MATRIX-ISOLATION INFRARED SPECTROMETRY
HIGH-RESOLUTION GAS-CHROMATOGRAPHY MATRIX-ISOLATION INFRARED SPECTROMETRY
复制标题
DOI:
10.1021/ac00285a024
复制
发表时间:
1985-01-01
影响因子:
7.4
通讯作者:
BOURNE, S
中科院分区:
文献类型:
--
作者:
REEDY, GT;ETTINGER, DG;BOURNE, S
An apparatus Is described that allows the collection, within a matrix of condensed Inert gas, of the effluent compounds from a hlgh-resolutlon gas chromatograph. Each collected compound Is contained within an area typically 0.3 mmIn diameter yielding a concomitant high level of Infrared spectral absorbance per nanogram of sample. Tests demonstrate the level of Infrared sensitivity, the achievement of matrix Isolation of the sample molecules, and the maintenance of compound separation achieved by the gas chromatograph. Use of the apparatus Is demonstrated for PAH, PCB, dioxin, and aliphatic hydrocarbon compounds.The technique of matrix isolation has beenextensively used to study the infrared spectrometric properties of molecules (1-4). In this technique themolecules to be studied are co-condensed with an excess of inert gas, such as argon. Trapped in the solid argon matrix, at temperatures between 10 and 15