CORRECTING FOR CELL TILT CAUSED BY DIFFERENTIAL EXPANSION OF ROTORS IN ANALYTICAL ULTRACENTRIFUGE

CORRECTING FOR CELL TILT CAUSED BY DIFFERENTIAL EXPANSION OF ROTORS IN ANALYTICAL ULTRACENTRIFUGE
复制标题

DOI:
10.1016/0003-2697(63)90107-6
复制
发表时间:
1963-01-01
影响因子:
2.9
通讯作者:
GROPPER, L
GROPPER, L
中科院分区:
生物学4区
文献类型:
--
作者:
GROPPER, L

文献摘要

被引文献

相似文献

由于施加在其单元孔顶部和底部的重量不同,分析转子以不同的方式膨胀。这种不同的膨胀会导致电池倾斜,使光线以与法线不同的角度进入溶液。描述了两种处理电池倾斜的过程:第一种是通过校正重量差异来消除电池倾斜;第二种是通过重新定位光源来补偿电池倾斜。
Because there is a difference in the weight applied against the top and bottom of its cell hole, an analytical rotor expands differentially. This differential expansion causes the cell to tilt so that light enters the solution at various angles to the normal. Two procedures for dealing with cell tilt are described: the first eliminates cell tilt by correcting the disparity in weight; the second compensates for cell tilt by repositioning the light source.