Uncertainty Quantification for Non-invasive Assessment of Pressure Drop Across a Coarctation of the Aorta Using CFD

Uncertainty Quantification for Non-invasive Assessment of Pressure Drop Across a Coarctation of the Aorta Using CFD
复制标题

DOI:
10.1007/s13239-018-00381-3
复制
发表时间:
2018-12-01
影响因子:
1.8
通讯作者:
Goubergrits, Leonid
Goubergrits, Leonid
中科院分区:
工程技术4区
文献类型:
--
作者:
Bruening, Jan;Hellmeier, Florian;Goubergrits, Leonid

文献摘要

被引文献

相似文献

目的——使用 CFD 对主动脉缩窄压降进行数值评估是替代侵入性导管测量的一种有前途的方法。本研究的目的是调查和量化医学图像处理的两个基本步骤中引入的压降数值计算的不确定性:患者特定几何形状的分割和通过 4D-flow-MRI 测量患者特定流速。方法 - 基于基线分割,为 10 名患者的样本生成具有不同狭窄直径的几何形状。使用计算流体动力学计算不同体积流量下这些几何形状产生的压降。基于这些模拟,计算了二阶多项式拟合。基于这些多项式拟合,对压降计算的不确定性进行了量化。结果-计算出的压降值在患者之间差异很大。在 4 名患者中,发现压力下降高于或低于 20mmHg 的临床阈值。压降的中位标准偏差为2.3mmHg。压降对体积流量或狭窄几何形状变化的敏感性因患者而异。结论-图像分割和体积流量测量过程中的不确定性引入的数值压降计算的不确定性与使用侵入性导管插入术的压降测量的不确定性相当。然而,对于一些患者来说,这种不确定性会导致不同的治疗决定。因此,患者特定的不确定性评估可能有助于更好地了解主动脉缩窄压降时数值计算的生物标志物的可靠性。
Purpose-Numerical assessment of the pressure drop across an aortic coarctation using CFD is a promising approach to replace invasive catheter-based measurements. The aim of this study was to investigate and quantify the uncertainty of numerical calculation of the pressure drop introduced during two essential steps of medical image processing: segmentation of the patient-specific geometry and measurement of patient-specific flow rates from 4D-flow-MRI.Methods-Based on the baseline segmentation, geometries with different stenosis diameters were generated for a sample of ten patients. The pressure drop generated by these geometries was calculated for different volume flow rates using computational fluid dynamics. Based on these simulations, a second order polynomial fit was calculated. Based on these polynomial fits an uncertainty of pressure drop calculation was quantified.Results-The calculated pressure drop values varied strongly between the patients. In four patients, pressure drops above and below the clinical threshold of 20mmHg were found. The median standard deviation of the pressure drop was 2.3mmHg. The sensitivity of the pressure drop toward changes in the volume flow rate or the stenosis geometry varied between patients.Conclusion-The uncertainty of numerical pressure drop calculation introduced by uncertainties during image segmentation and measurement of volume flow rates was comparable to the uncertainty of pressure drop measurements using invasive catheterization. However, in some patients this uncertainty would have led to different treatment decision. Therefore, patient-specific uncertainty assessment might help to better understand the reliability of a numerically calculated biomarker as the pressure drop across an aortic coarctation.