Quantitative computed tomography for spinal density measurement. Factors affecting precision.

Quantitative computed tomography for spinal density measurement. Factors affecting precision.
复制标题

用于脊柱密度测量的定量计算机断层扫描。

DOI:
--
复制
发表时间:
1985
影响因子:
6.7
通讯作者:
R. Neer
R. Neer
中科院分区:
医学1区
文献类型:
--
作者:
D. Rosenthal;M. Ganott;G. Wyshak;D. Slovik;S. Doppelt;R. Neer

文献摘要

被引文献

相似文献

Quantitative computed tomography (QCT) was performed in duplicate on 84 patients to test the short-term precision of the technique. Statistical analysis of the data revealed that precision was not a function of spinal density. It appeared to be worse in osteopenic individuals only when expressed as a percentage. Precision was slightly better in male than in female patients. There is a 90% likelihood that a duplicate measurement will fall within 20 CT units of the first determination in female patients and within ten units in male patients.