Measurement of the electrically evoked compound action potential via a neural response telemetry system

Measurement of the electrically evoked compound action potential via a neural response telemetry system
复制标题

DOI:
10.1177/000348940211100505
复制
发表时间:
2002-05-01
影响因子:
1.4
通讯作者:
von Wallenberg, E
von Wallenberg, E
中科院分区:
医学3区
文献类型:
--
作者:
Dillier, N;Lai, WK;von Wallenberg, E

文献摘要

被引文献

相似文献

本研究的主要目的是验证一种新的技术--神经反应遥测技术(NRT)--用于测量成年人工耳蜗者通过其Nucleus CI24M植入物的电诱发复合动作电位。使用NRT软件对38名成年人进行了各种测量程序的评估。电诱发复合动作电位在38例成人中有31例(81.6%),在160个电极中有132例(82.5%)。除了验证这项技术,我们还建立了一组默认的临床测试参数。
The main aim of this study was to validate a new technique, neural response telemetry (NRT), for measuring the electrically evoked compound action potential in adult cochlear implant users via their Nucleus CI24M implant. Thirty-eight adults were evaluated with a variety of measurement procedures with the NRT software. Electrically evoked compound action potentials were obtained in 31 of the 38 adults (81.6%) and in 132 of the 160 electrodes (82.5%) tested. In addition to validating this technique, We also established a set of default clinical test parameters.