Simultaneous PET and MR imaging
Simultaneous PET and MR imaging
复制标题
DOI:
10.1088/0031-9155/42/10/010
复制
发表时间:
1997-10-01
影响因子:
3.5
通讯作者:
Marsden, PK
中科院分区:
文献类型:
--
作者:
Shao, YP;Cherry, SR;Marsden, PK
We have developed a prototype PET detector which is compatible with a clinical MRI system to provide simultaneous PET and MR imaging. This single-slice PET system consists of 48 2 x 2 x 10 mm(3) LSO crystals in a 38 mm diameter ring configuration that can be placed inside the receiver coil of the MRI system, coupled to three multi-channel photomultipliers housed outside the main magnetic field via 4 m long and 2 mm diameter optical fibres. The PET system exhibits 2 mm spatial resolution, 41% energy resolution at 511 keV and 20 ns timing resolution. Simultaneous PET and MR phantom images were successfully acquired.