Readout of X-ray Pulses from a Single-pixel TES Microcalorimeter with Microwave Multiplexer Based on SQUIDs Directly Coupled to Resonators

Readout of X-ray Pulses from a Single-pixel TES Microcalorimeter with Microwave Multiplexer Based on SQUIDs Directly Coupled to Resonators
复制标题

使用基于直接耦合到谐振器的 SQUID 的微波多路复用器从单像素 TES 微量热计读取 X 射线脉冲

DOI:
10.1007/s10909-018-2030-0
复制
发表时间:
2018
影响因子:
2
通讯作者:
Mitsuda K.
Mitsuda K.
中科院分区:
物理与天体物理3区
文献类型:
--
作者:
Nakashima Y.;Hirayama F.;Kohjiro S.;Yamamori H.;Nagasawa S.;Sato A.;Irimatsugawa T.;Muramatsu H.;Hayashi T.;Yamasaki N. Y.;Mitsuda K.

文献摘要

相似文献

我们首先展示了X射线TES和直接耦合微波SQUID复用器(D-coup)之间的合作。MW-Mux),其中SQUID直接耦合到谐振器。使用D-coup。MW-Mux,我们成功地检测到来自单像素TES的X射线脉冲,并评估了5.9 keV下11.8 eV FWHM的能量分辨率。我们发现这种初步的能量分辨率归因于目前TES的非最佳偏置点,这可以通过未来的读出噪声的降低来改善。
We have first demonstrated the cooperation between an X-ray TES and a directly coupled microwave SQUID multiplexer (D-coup. MW-Mux) in which SQUIDs are directly coupled to resonators. Using D-coup. MW-Mux, we successfully detected the X-ray pulses from a single-pixel TES and evaluated the energy resolution of 11.8 eV FWHM at 5.9 keV. We found this preliminary energy resolution is attributed to non-optimum bias point of the present TES, which can be improved by the future reduction of readout noise.