10 MV x-ray zero-area phantom scatter correction factors (Sp) obtained using three extrapolation methods.
10 MV x-ray zero-area phantom scatter correction factors (Sp) obtained using three extrapolation methods.
复制标题
使用三种外推方法获得的 10 MV X 射线零面积体模散射校正因子 (Sp)。
DOI:
--
复制
发表时间:
1996
影响因子:
3.5
通讯作者:
A. Iwasaki
中科院分区:
文献类型:
--
作者:
A. Iwasaki
Three sets of 10 MV x-ray zero-area phantom scatter correction factors (Sp) have been obtained using three methods. The three methods are all based on the Bjärngard-Petti extrapolation principle. The three sets of data assume lateral CPE (charged particle equilibrium) for the primary absorbed dose. Using the most reliable set of data, a set of 10 MV x-ray SMRs (scatter-maximum ratios) is produced and parameterized. With respect to the zero-area Sp correction factor at a depth of 2.5 cm, the parameterized expression gives Sp (0; 2.5) = 0.931 and a Rice-Chin Monte Carlo simulation gives Sp (0; 2.5) = 0.927. The former is only 0.4% larger than the latter.
影响因子:
3.5
作者:
Rice,RK;Chin,LM
通讯作者:
Chin,LM