10 MV x-ray zero-area phantom scatter correction factors (Sp) obtained using three extrapolation methods.

10 MV x-ray zero-area phantom scatter correction factors (Sp) obtained using three extrapolation methods.
复制标题

使用三种外推方法获得的 10 MV X 射线零面积体模散射校正因子 (Sp)。

DOI:
--
复制
发表时间:
1996
影响因子:
3.5
通讯作者:
A. Iwasaki
A. Iwasaki
中科院分区:
工程技术2区
文献类型:
--
作者:
A. Iwasaki

文献摘要

参考文献

被引文献

相似文献

三套10 MV的X射线零面积体模散射校正因子(Sp)已获得使用三种方法。这三种方法都是基于Bjärngard-Petti外推原理。这三组数据假定初级吸收剂量的横向CPE(带电粒子平衡)。使用最可靠的数据集,一组10 MV的X射线SMR(散射最大值比)的生产和参数化。对于2.5 cm深度处的零面积Sp校正因子,参数化表达式给出Sp(0; 2.5)= 0.931,Rice-Chin Monte Carlo模拟给出Sp(0; 2.5)= 0.927。前者仅比后者大0.4%。
Three sets of 10 MV x-ray zero-area phantom scatter correction factors (Sp) have been obtained using three methods. The three methods are all based on the Bjärngard-Petti extrapolation principle. The three sets of data assume lateral CPE (charged particle equilibrium) for the primary absorbed dose. Using the most reliable set of data, a set of 10 MV x-ray SMRs (scatter-maximum ratios) is produced and parameterized. With respect to the zero-area Sp correction factor at a depth of 2.5 cm, the parameterized expression gives Sp (0; 2.5) = 0.931 and a Rice-Chin Monte Carlo simulation gives Sp (0; 2.5) = 0.927. The former is only 0.4% larger than the latter.
蒙特卡罗计算散射与初级比率,以实现初级和散射剂量的标准化。
DOI: 10.1088/0031-9155/35/3/003
发表时间: 1990
影响因子: 3.5
作者:
Rice,RK;Chin,LM
通讯作者: Chin,LM