Repeatability and reproducibility of high concentration data in reversed-phase liquid chromatography. I. Overloaded band profiles on Kromasil-C18.
Repeatability and reproducibility of high concentration data in reversed-phase liquid chromatography. I. Overloaded band profiles on Kromasil-C18.
复制标题
反相液相色谱中高浓度数据的重复性和再现性。
DOI:
10.1016/s0021-9673(03)00685-x
复制
发表时间:
2003
期刊:
影响因子:
--
通讯作者:
G. Guiochon
中科院分区:
文献类型:
--
作者:
F. Gritti;G. Guiochon
Single-component adsorption-isotherm data were acquired by frontal analysis (FA) for six low-molecular-mass compounds (phenol, aniline, caffeine, theophylline, ethylbenzene and propranolol) on one Kromasil-C18column, using water–methanol solutions (between 70:30 and 20:80, v/v) as the mobile phase. Propranolol data were also acquired using an acetate buffer (0.2 M) instead of water. The data were modeled for best agreement between calculated and experimental overloaded band profiles. The adsorption energy distribution was also derived and used for the selection of the best isotherm model. Widely different isotherm models were found to model best the data obtained for these compounds, convex upward (i.e. Langmuirian), convex downward (i.e. anti-Langmuirian), and S-shaped isotherms. Using the same sample size for all columns (loading factor, Lf≈10%), overloaded band profiles were recorded on four different columns packed with the same batch of Kromasil-C18and five other columns packed with different batches of Kromasil-C18. These experimental band profiles were compared to the profile calculated from the isotherm measured by FA on the first column. The repeatability as well as the column-to-column and the batch-to-batch reproducibilities of the band profiles are better than 4%.
影响因子:
7.4
作者:
Umpleby, RJ;Baxter, SC;Shimizu, KD
通讯作者:
Shimizu, KD