Imaging characteristics of a high resolution cone beam collimator
Imaging characteristics of a high resolution cone beam collimator
复制标题
高分辨率锥形束准直器的成像特性
DOI:
10.1109/23.12804
复制
发表时间:
1988
影响因子:
1.8
通讯作者:
R. Coleman
中科院分区:
文献类型:
--
作者:
R. Jaszczak;K. Greer;C. Floyd;S. Manglos;R. Coleman
Single-photon-emission computed tomography (SPECT) using cone-beam geometry requires accurately built collimators and an appropriate reconstruction algorithm. The authors have investigated the imaging characteristics of a cast, high-resolution, cone-beam collimator and have compared its measured sensitivity and resolution with analytically predicted values. Using a filtered back-projection algorithm, they have measured point-spread functions at several locations within the field of view. Measurements have been made both with the collimator surface parallel to the axis-of-rotation (AOR) and with the surface tilted at an angle of 20 degrees with respect to the AOR. Although the SPECT point-spread response is not completely shift-invariant, high-quality cone-beam images of a brain phantom have been obtained. The measured performance of the collimator agrees well with analytically derived results. >