All-optical scanhead for ultrasound and photoacoustic imaging-Imaging mode switching by dichroic filtering.
All-optical scanhead for ultrasound and photoacoustic imaging-Imaging mode switching by dichroic filtering.
复制标题
DOI:
10.1016/j.pacs.2013.12.002
复制
发表时间:
2014-03
期刊:
影响因子:
7.9
通讯作者:
Li, Pai-Chi
中科院分区:
文献类型:
--
作者:
Hsieh, Bao-Yu;Chen, Sung-Liang;Ling, Tao;Guo, L Jay;Li, Pai-Chi
Ultrasound (US) and photoacoustic (PA) multimodality imaging has the advantage of combining good acoustic resolution with high optical contrast. The use of an all-optical scanhead for both imaging modalities can simplify integration of the two systems and miniaturize the imaging scanhead. Herein we propose and demonstrate an all-optical US/PA scanhead using a thin plate for optoacoustic generation in US imaging, a polymer microring resonator for acoustic detection, and a dichroic filter to switch between the two imaging modes by changing the laser wavelength. A synthetic-aperture focusing technique is used to improve the resolution and contrast. Phantom images demonstrate the feasibility of this design, and show that axial and lateral resolutions of 125 μm and 2.52°, respectively, are possible.