XAFS Analysis of Ti and Ni Dissolution from Pure Ti, Ni-Ti Alloy, and SUS304 in Soft Tissues
XAFS Analysis of Ti and Ni Dissolution from Pure Ti, Ni-Ti Alloy, and SUS304 in Soft Tissues
复制标题
DOI:
10.1246/cl.2008.958
复制
发表时间:
2008-09-05
影响因子:
1.6
通讯作者:
Watari, Fumio
中科院分区:
文献类型:
--
作者:
Uo, Motohiro;Asakura, Kiyotaka;Watari, Fumio
Ti, a Ni-Ti alloy, and stainless steel (SUS304) were implanted in rat soft tissues for 6 months. The chemical states of the component elements contained in the tissues surrounding those metallic implants were analyzed by fluorescence X-ray absorption line structure (XAFS). Titanium oxide and a nickel aquo complex, which Would be Caused by the erosion of Ti and Ni, were detected in the Surrounding tissues. The concentrations of Ti and Ni in the tissues were 1-3 and 300 ppm, respectively. Thus, the dissolved conditions and concentrations of those practical metallic materials in vivo were successfully confirmed.