Energy- resolved Computed Tomography Measurements of Iron Solution and Adipose as a Simulation for Estimating the Iron Concentration in the Human Liver
Energy- resolved Computed Tomography Measurements of Iron Solution and Adipose as a Simulation for Estimating the Iron Concentration in the Human Liver
复制标题
铁溶液和脂肪的能量分辨计算机断层扫描测量作为估计人类肝脏中铁浓度的模拟
DOI:
10.1080/00223131.2013.773163
复制
发表时间:
2013
影响因子:
1.2
通讯作者:
I. Kanno, 他.
中科院分区:
文献类型:
--
作者:
Y.Yamashita;I. Kanno, 他.
To measure the iron concentration in the human liver, which comprises soft tissue and adipose as well as iron, energy-resolved computed tomography (CT) is applied using a transXend detector. The transXend detector measures X-rays as electric current and gives an energy distribution after analysis. Energy-resolved CT clearly separates iron solution from acrylic, a substitute for soft tissue, whereas conventional current-measurement CT cannot. Using CT values obtained in two different energy ranges, a two-dimensional map of iron, adipose and soft tissue is plotted. With this map, the components in the liver can be identified.