Energy- resolved Computed Tomography Measurements of Iron Solution and Adipose as a Simulation for Estimating the Iron Concentration in the Human Liver

Energy- resolved Computed Tomography Measurements of Iron Solution and Adipose as a Simulation for Estimating the Iron Concentration in the Human Liver
复制标题

铁溶液和脂肪的能量分辨计算机断层扫描测量作为估计人类肝脏中铁浓度的模拟

DOI:
10.1080/00223131.2013.773163
复制
发表时间:
2013
影响因子:
1.2
通讯作者:
I. Kanno, 他.
I. Kanno, 他.
中科院分区:
工程技术4区
文献类型:
--
作者:
Y.Yamashita;I. Kanno, 他.

文献摘要

相似文献

为了测量人类肝脏中的铁浓度,其包括软组织和脂肪以及铁,使用transXend检测器应用能量分辨计算机断层扫描(CT)。transXend探测器将X射线测量为电流,并在分析后给出能量分布。能量分辨CT可以清楚地将铁溶液与丙烯酸树脂(软组织的替代品)分开,而传统的电流测量CT则不能。使用在两个不同能量范围内获得的CT值,绘制铁、脂肪和软组织的二维图。有了这个地图,可以识别肝脏中的成分。
To measure the iron concentration in the human liver, which comprises soft tissue and adipose as well as iron, energy-resolved computed tomography (CT) is applied using a transXend detector. The transXend detector measures X-rays as electric current and gives an energy distribution after analysis. Energy-resolved CT clearly separates iron solution from acrylic, a substitute for soft tissue, whereas conventional current-measurement CT cannot. Using CT values obtained in two different energy ranges, a two-dimensional map of iron, adipose and soft tissue is plotted. With this map, the components in the liver can be identified.