The Clinical Impact of Incomplete Left Atrial Appendage Closure With the Watchman Device in Patients With Atrial Fibrillation
The Clinical Impact of Incomplete Left Atrial Appendage Closure With the Watchman Device in Patients With Atrial Fibrillation
复制标题
DOI:
10.1016/j.jacc.2011.11.028
复制
发表时间:
2012-03-06
影响因子:
24
通讯作者:
Reddy, Vivek Y.
中科院分区:
文献类型:
--
作者:
Viles-Gonzalez, Juan F.;Kar, Saibal;Reddy, Vivek Y.
Objectives The purpose of this study was to investigate the frequency and clinical impact of incomplete left atrial appendage (LAA) sealing and consequent peri-device residual blood flow in patients undergoing percutaneous LAA closure with the Watchman device (Atritech, Inc., Plymouth, Minnesota).Background During percutaneous LAA closure for stroke prophylaxis, the geometric variability of the LAA ostium may result in an incomplete seal of the LAA. On the one hand, this could enhance thrombus formation and embolization of thrombi around the device into the circulation; on the other hand, the relatively small size of these leaks may preclude clinically relevant embolizations.Methods Patients randomly assigned to device implantation in the PROTECT AF (Percutaneous Closure of the Left Atrial Appendage Versus Warfarin Therapy for Prevention of Stroke in Patients With Atrial Fibrillation) trial were analyzed. Transesophageal echocardiography was performed at 45 days, 6 months, and 12 months. Per the study protocol, patients discontinued warfarin therapy if the 45-day Transesophageal echocardiogram revealed either minimal or no peri-device flow (jet