Performance of the helium dewar and the cryocoolers of the Hitomi soft x-ray spectrometer
Performance of the helium dewar and the cryocoolers of the Hitomi soft x-ray spectrometer
复制标题
Hitomi 软 X 射线光谱仪的氦杜瓦瓶和制冷机的性能
DOI:
10.1117/1.jatis.4.1.011208
复制
发表时间:
2017
期刊:
影响因子:
--
通讯作者:
Hitomi collaboration incl. Kasuke Sato
中科院分区:
文献类型:
--
作者:
R. Farwig;Y. Giga and P.-Y. Hsu;Hitomi collaboration incl. Kasuke Sato
The soft x-ray spectrometer (SXS) was a cryogenic high-resolution x-ray spectrometer onboard the Hitomi (ASTRO-H) satellite that achieved energy resolution of 5 eV at 6 keV, by operating the detector array at 50 mK using an adiabatic demagnetization refrigerator (ADR). The cooling chain from room temperature to the ADR heat sink was composed of two-stage Stirling cryocoolers, aJoule–Thomson cryocooler, and superfluid liquid helium and was installed in a dewar. It was designed to achieve a helium lifetime of more than 3 years with a minimum of 30 L. The satellite was launched on February 17, 2016, and the SXS worked perfectly in orbit, until March 26 when the satellite lost its function. It was demonstrated that the heat load on the helium tank was about 0.7 mW, which would have satisfied the lifetime requirement. This paper describes the design, results of ground performance tests, prelaunch operations, and initial operation and performance in orbit of the flight dewar and the cryocoolers.