Recessed oxygen electrodes: getting more than PO2.
Recessed oxygen electrodes: getting more than PO2.
复制标题
嵌入式氧电极:获取超过 PO2。
DOI:
10.1007/978-1-4615-0205-0_29
复制
发表时间:
2003
影响因子:
--
通讯作者:
Buerk,DonaldG
中科院分区:
文献类型:
--
作者:
Buerk,DonaldG
Several different experimental techniques have been developed using recessed gold cathode (Whalen/Nair-type) 02microelectrodes to investigate 02transport to tissue. After relatively large 02cathodes were first used in tissue over 60 years ago, serious measurement errors were soon recognized, including electrochemical depletion of tissue 02, sensitivity to blood flow or convective mixing (stirring artifact), signal averaging due to the tip size, and damage to cells or the microcirculation during penetration. To address these problems, 02electrodes were miniaturized, although this did not always eliminate PO2measurement errors.