A comparison of response characteristics of airflow and pressure transducers commonly used in rhinomanometry.
A comparison of response characteristics of airflow and pressure transducers commonly used in rhinomanometry.
复制标题
鼻测压中常用的气流和压力传感器的响应特性比较。
DOI:
10.1109/10.102806
复制
发表时间:
1990
期刊:
影响因子:
--
通讯作者:
Doty,RL
中科院分区:
文献类型:
--
作者:
Frye,RE;Doty,RL
Although the ranges in which pneumotachographs evidence linear output to static flows are documented in the literature, output reliability or the stability of calibration functions resulting from the input of dynamic nonsinusoidal flows (such as those which occur during nasal breathing) have not been investigated. Furthermore, it is not known whether the type of requisite pressure transducer used in conjunction with the pneumotachograph influences the pneumotachograph's linearity, output reliability, or dynamic response. To provide information on these points, the authors determined the dynamic and static responsiveness of three pneumotachographs commonly used in rhinomanometry, in combination with three requisite pressure transducers. Experimental data indicate that: (a) output reliability depended on the pneumotachograph-pressure transducer combination and was not readily predictable from the reliabilities of the individual components, (b) heating increased pneumotachograph reliability, and (c) differences in accuracy existed among transducer combinations of high, but not low, flow frequencies.< >