Indirect determination of paracetamol in pharmaceutical formulations by inhibition of the system luminol-H2O2-Fe(CN)3-(6) chemiluminescence.
Indirect determination of paracetamol in pharmaceutical formulations by inhibition of the system luminol-H2O2-Fe(CN)3-(6) chemiluminescence.
复制标题
通过抑制鲁米诺-H2O2-Fe(CN)3-(6) 化学发光系统来间接测定药物制剂中的扑热息痛。
DOI:
10.1016/s0731-7085(99)00124-7
复制
发表时间:
1999
影响因子:
3.4
通讯作者:
J. M. Calatayud
中科院分区:
文献类型:
--
作者:
A. Alapont;L. Zamora;J. M. Calatayud
After a large drug scanning, the system Luminol–H2O2–Fe(CN)63−is proposed for first time for the indirect determination of paracetamol. The method is based on the oxidation of paracetamol by hexacyanoferrate (III) and the subsequent inhibitory effect on the reaction between luminol and hydrogen peroxide. The procedure resulted in a linear calibration graph over the range 2.5–12.5 μg ml−1of paracetamol with a sample throughput of 87 samples h−1. The influence of foreign compounds was studied and, the method was applied to determination of the drug in three different pharmaceutical formulations.