Continuous measurement of minute ventilation and gaseous metabolism of newborn infants.
Continuous measurement of minute ventilation and gaseous metabolism of newborn infants.
复制标题
连续测量新生儿每分钟通气量和气体代谢。
DOI:
10.1152/jappl.1981.50.5.1098
复制
发表时间:
1981
期刊:
影响因子:
--
通讯作者:
James,LS
中科院分区:
文献类型:
--
作者:
Schulze,K;Kairam,R;Stefanski,M;Sciacca,R;James,LS
A system of instrumentation for continuous measurement of gaseous metabolism and minute volume (VI) in the human newborn is described. O2 uptake and CO2 production are measured by open-circuit techniques utilizing a Servomex OA184 differential paramagnetic O2 analyzer and a BEckman LB-2 infrared CO2 analyzer. VI is measured with bias-flow pneumotachometry. Bench performance is described, methodological errors are defined, and clinical data are presented. The instrumentation is capable of safe, accurate, and continuous measurement of respiratory and metabolic variables in low-birth-weight infants.