X-ray pulsing methods for reduced-dose computed tomography in PET/CT attenuation correction
X-ray pulsing methods for reduced-dose computed tomography in PET/CT attenuation correction
复制标题
PET/CT 衰减校正中减少剂量计算机断层扫描的 X 射线脉冲方法
DOI:
--
复制
发表时间:
2014
期刊:
影响因子:
--
通讯作者:
B. De Man
中科院分区:
文献类型:
--
作者:
U. Wiedmann;V. Neculaes;D. Harrison;E. Asma;Paul Kinahan;B. De Man
The image quality needed for CT-based attenuation correction (CTAC) is significantly lower than what is used currently for diagnostic CT imaging. Consequently, the X-ray dose required for sufficient image quality with CTAC is relatively small, potentially smaller than the lowest X-ray dose clinical CT scanners can provide. Operating modes have been proposed in which the X-rays are periodically turned on and off during the scan in order to reduce X-ray dose. This study reviews the different methods by which X-rays can be modulated in a CT scanner, and assesses their adequacy for lowdose acquisitions as required for CTAC. Calculations and experimental data are provided to exemplify selected X-ray pulsing scenarios. Our analysis shows that low-dose pulsing is possible but challenging with clinically available CT tubes. Alternative X-ray tube designs would lift this restriction.